Z-0064-2015 Class III Terminated
FDA device recall Z-0064-2015 was initiated by Roche Diagnostics Operations, Inc. on Sep 12, 2014 and is designated Class III. Reason for recall: The default references for normal values are inconsistent between cobas b 221 and other blood gas instruments. In addition, the reference ranges are different when comparing with corresponding ranges… The recall status is terminated (terminated Jan 10, 2018). Affected quantity: 492 devicees.
Recall Details
- Product Type
- Devices
- Report Date
- Oct 22, 2014
- Initiation Date
- Sep 12, 2014
- Termination Date
- Jan 10, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 492 devicees
Product Description
Omni Micro-electrode/reference electrode for cobas b221 analyzer. The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
Reason for Recall
The default references for normal values are inconsistent between cobas b 221 and other blood gas instruments. In addition, the reference ranges are different when comparing with corresponding ranges listed in the Instructions for Use for the analyzer and the reference literature source.
Distribution Pattern
Nationwide Distribution including Puerto Rico and Guam, and the states of AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, ME, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA and WI.
Code Information
Model/Catalog/Part Number: 1. 03337103001 2. 03337111001 3. 03337154001 4. 03337138001 5. 03337120001 6. 03337146001 7. 03337111692 8. 03337138692 9. 03337154692