Z-0065-2015 Class II Terminated

Recalled by Stryker Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-0065-2015 was initiated by Stryker Howmedica Osteonics Corp. on Jul 24, 2014 and is designated Class II. Reason for recall: Stryker Orthopaedics received a report indicating some of the Nail Insertion Sleeves had a different inner diameter than described on the package label. The recall status is terminated (terminated Jun 8, 2017). Affected quantity: 484 units.

Recall details

Product Type
Devices
Report Date
Oct 22, 2014
Initiation Date
Jul 24, 2014
Termination Date
Jun 8, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
484 units

Product description

Stryker T2 Tibia System Nail Insertion Sleeve, Elastic, Sterile, a) REF 1806-1406S, SPI 08-11 and b) REF 1806-1407S, SPI 08-13. Product Usage: The T2 Tibial Nailing system is intended to provide temporary stabilization of various types of fractures, malunion and nonunion of the tibia. The nails are inserted using an opened or closed technique and can be statically, dynamically and compressed locked. The system is indicated for long bone fracture fixation, specifically tibial fracture fixation. The sleeves are intended to ease nail insertion and to protect the condyle surfaces during nail insertion.

Reason for recall

Stryker Orthopaedics received a report indicating some of the Nail Insertion Sleeves had a different inner diameter than described on the package label.

Distribution pattern

US Nationwide Distribution in the states of CA, FL, IN, MD, NC, NH, NY, OH, PA and WI.

Code information

Catalog #1806-1406S - Lot #'s K019A3C, K031467, K032811, K032812, K06AAA8, K07F95F and K083672, expiry date 3/31/2019 and Catalog #1806-1407S - Lot #'s K031468, K03E8AD and K074529, expiry date 3/31/2019

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0065-2015
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0065-2015. Report a data issue: hello@regdatalab.com.