Z-0070-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
November 15, 2017
Initiation Date
April 25, 2017
Termination Date
February 6, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product Description

Custom Ti MAK RS OSS Segmental Femoral Rotating Hinged Knee. This Rotating Hinged Knee is intended to replace patient anatomy.

Reason for Recall

A custom implant manufactured entirely from titanium for a patient with a nickel allergy was requested by a surgeon. After the device was manufactured and used in the planned surgery, it was discovered internally that the implant assembly had a CoCrMo screw included as a component of the implant, which contains nickel.

Distribution Pattern

This was a custom implant, there was only one (1) unit manufactured and distributed. The product has been implanted, so it is not subject to being returned.

Code Information

Part Number: CP116574 Lot Number: 073840