Z-0071-2017 Class II Terminated
Recalled by GE OEC Medical Systems, Inc
— Salt Lake City, UT
FDA device recall Z-0071-2017 was initiated by GE OEC Medical Systems, Inc on Sep 23, 2016 and is designated Class II. Reason for recall: GE Healthcare Surgery announces a voluntary field action for the OEC Brivo 715 Prime, OEC Brivo 685 Essential, and OEC Brivo 865 Advance because the circuit board that controls vertical motion of the… The recall status is terminated (terminated Apr 24, 2017). Affected quantity: 148 total.
Recall details
- Product Type
- Devices
- Report Date
- Oct 19, 2016
- Initiation Date
- Sep 23, 2016
- Termination Date
- Apr 24, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 148 total
Product description
OEC Brivo 715 Prime, OEC Brivo 785 Essential, OEC Brivo 865 Advance Product Usage: The OEC Brivo Mobile C-Arm X-Ray Products are designed to provide digital spot film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments. The product is used for general surgical applications and musculoskeletal procedures to visualize, for example, implant localizations or needle positions for aspirations, injections or biopsy. The OEC Brivo is not indicated for interventional use.
Reason for recall
GE Healthcare Surgery announces a voluntary field action for the OEC Brivo 715 Prime, OEC Brivo 685 Essential, and OEC Brivo 865 Advance because the circuit board that controls vertical motion of the C-arm may fail, resulting in the C-Arm moving up or down without command.
Distribution pattern
Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, FL, KS, MS, NJ, NV, NY, PA, TX, UT, and WV. No Candaian or Va/govt/military consignees. There are 133 other foreign consignees.
Code information
GE Brivo OEC715: Serial Numbers B2S15158 to B2S15164; B2S15166 to B2S15171; B2S15173 to B2S15195; B2S15197; B2S15200; B2S15202; B2S16046; B2S16050; B2S16051 GE Brivo OEC785: Serial Numbers B3S15236; B3S15241; B3S15242; B3S15246; B3S15248; B3S15250; B3S15253 to B3S15257; B3S15259 to B3S15338; B3S15353; B3S15357 to B3S15359; B3S16152 GE Brivo OEC865: Serial Numbers B4S15063 to B4S15070; B4S15072; B4S15073; B4S15075; B4S16029
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0071-2017
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for GE OEC Medical Systems, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works