Z-0071-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 11, 2023
- Initiation Date
- August 23, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 256 units
Product Description
LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Arm System Model Numbers LFSST2323, LFSLFSST29TV23, LFSST29TV23, LFSST23TV, LFSST23, LFSST2323TV, LFSLFSST2323, LDFSLFSLED7TV5, LDFSLFSLED7TV55, LFSLED55, LFSLFSST23, LFSST2923TV, LFSST29TV, LFSST29, LFSST19WC, LFSST2923, LFSLFSST-1, LFSST292323, LFSST1919WC, LFS ST2319WC
Reason for Recall
The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.
Distribution Pattern
US States: MI
Code Information
Serial Numbers: 06087323 - 06087352, 06087353 - 06087354, 06081292 - 06081293, 06087448 - 06087449, 06087450 - 06087469, 06097595 - 06097596, 06107716 - 06107718, 06107719 - 06107728, 06107730, 06107731 - 06107732, 06111348 - 06111349, 06117855 - 06117856, 06117857 - 06117866, 06117868 - 06117868, 06111350 - 06111351, 06117949 - 06117949, 06117954 - 06117973, 06118031 - 06118032, 06128116 - 06128131, 06128172 - 06128176, 06128177 - 06128186, 06128187 - 06128188, 07010004, 07010002, 07010005 - 07010006, 07028436 - 07028441, 07030080 - 07030081, 07048491 - 07048495, 07058506 - 07058507, 07058508 - 07058512, 07068602 - 07068604, 07068605 - 07068607, 07068611 - 07068622, 07068609 - 07068610, 07068667 - 07068670, 07061479 - 07061484, 07078679 - 07078680, 07071485 - 07071488, 07071501, 07088741 - 07088742, 07081124, 07088766, 07098769 - 07098773, 07098767 - 07098768, 07098774 - 07098775, 07108885 - 07108889, 07108883 - 07108884, 07108876 - 07108882, 07108890 - 07108891, 07101533 - 07101534, 07129027 - 07129028, RA000182, RA000184, 08029174 - 08029183, 08029187 - 08029188, 08029189 - 08029190, 09111694, 10011776, 10011777, 10011778, 10011779, 10011780, 10011781, 10011782, 10051792, 10051794, 10051796, 10051798, 10051800, 10051802, 10051804, 100610044, 100610045, 100610046, 100610047, 100610048, 100610049, 100610050, 100610051, 100610052, 100610053, 101010180, 101010182, 101010184, 101010186, 101010188, 101010190, 101010192, 101210252, BM005435 - BM005435, BM006121 - BM006121, BM008121 - BM008127