Z-0071-2027 Class II Ongoing

Recalled by Steris Corporation — Mentor, OH

FDA device recall Z-0071-2027 was initiated by Steris Corporation on Aug 31, 2026 and is designated Class II. Reason for recall: Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the… The recall status is ongoing. Affected quantity: 409.

Recall Details

Product Type
Devices
Report Date
Oct 14, 2026
Initiation Date
Aug 31, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
409

Product Description

VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007

Reason for Recall

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Distribution Pattern

Worldwide - US Nationwide distribution.

Code Information

UDI: 10724995220331 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0071-2027
openFDA file date
Oct 19, 2026
Loaded by RegDataLab
Oct 21, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.