Z-0073-2027 Class II Ongoing
FDA device recall Z-0073-2027 was initiated by Steris Corporation on Aug 31, 2026 and is designated Class II. Reason for recall: Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the… The recall status is ongoing. Affected quantity: 2205.
Recall Details
- Product Type
- Devices
- Report Date
- Oct 14, 2026
- Initiation Date
- Aug 31, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2205
Product Description
VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061
Reason for Recall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Distribution Pattern
Worldwide - US Nationwide distribution.
Code Information
UDI: 50724995023824 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027