Z-0154-2027 Class II Ongoing
FDA device recall Z-0154-2027 was initiated by Steris Corporation on Sep 21, 2026 and is designated Class II. Reason for recall: If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the… The recall status is ongoing. Affected quantity: 374 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 21, 2026
- Initiation Date
- Sep 21, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 374 units
Product description
RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115
Reason for recall
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
Distribution pattern
Worldwide - US Nationwide distribution.
Code information
UDI: 00724995245658 Lot numbers with expiration dates: 687046 10/5/2027 688017 10/19/2027