Z-0155-2027 Class II Ongoing

Recalled by Steris Corporation — Mentor, OH

FDA device recall Z-0155-2027 was initiated by Steris Corporation on Sep 21, 2026 and is designated Class II. Reason for recall: If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the… The recall status is ongoing. Affected quantity: 11806 units.

Recall details

Product Type
Devices
Report Date
Oct 21, 2026
Initiation Date
Sep 21, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11806 units

Product description

RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116

Reason for recall

If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.

Distribution pattern

Worldwide - US Nationwide distribution.

Code information

UDI: 00724995245955 Lot numbers with expiration dates: 689171 5/4/2027 689981 5/12/2027 690993 5/26/2027 691739 6/2/2027 692570 6/11/2027 694357 7/1/2027

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0155-2027
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0155-2027. Report a data issue: hello@regdatalab.com.