Z-0909-2026 Class II Ongoing

Recalled by Steris Corporation — Mentor, OH

Recall Details

Product Type
Devices
Report Date
December 17, 2025
Initiation Date
November 19, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10

Product Description

AMSCO 7052HP Washer/Disinfector Model/Catalog Number: 7052HP Product Description: The AMSCO 7052HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.

Reason for Recall

Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

Distribution Pattern

Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia.

Code Information

Lot Code: UDI:00724995176587, 00724995176594, 00724995176570 Serial number of distributed units: 3619625011, 3621625009, 3621625012, 3621925006, 3621925009, 3622425006, 3622725007, 3623425005, 3623825002, 3624525002