Z-0076-2022 Class II Ongoing
FDA device recall Z-0076-2022 was initiated by Datascope Corp. on Sep 8, 2021 and is designated Class II. Reason for recall: The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference. The recall status is ongoing. Affected quantity: 5054 total (US); 5033 total (OUS).
Recall details
- Product Type
- Devices
- Report Date
- Oct 13, 2021
- Initiation Date
- Sep 8, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5054 total (US); 5033 total (OUS)
Product description
Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Reason for recall
The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference.
Distribution pattern
Worldwide distribution - US Nationwide distribution.
Code information
All lots
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0076-2022
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026