Z-0076-2022 Class II Ongoing

Recalled by Datascope Corp. — Mahwah, NJ

Recall Details

Product Type
Devices
Report Date
October 13, 2021
Initiation Date
September 8, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5054 total (US); 5033 total (OUS)

Product Description

Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Reason for Recall

The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference.

Distribution Pattern

Worldwide distribution - US Nationwide distribution.

Code Information

All lots