Z-0076-2022 Class II Ongoing

Recalled by Datascope Corp. — Mahwah, NJ

FDA device recall Z-0076-2022 was initiated by Datascope Corp. on Sep 8, 2021 and is designated Class II. Reason for recall: The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference. The recall status is ongoing. Affected quantity: 5054 total (US); 5033 total (OUS).

Recall details

Product Type
Devices
Report Date
Oct 13, 2021
Initiation Date
Sep 8, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5054 total (US); 5033 total (OUS)

Product description

Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Reason for recall

The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference.

Distribution pattern

Worldwide distribution - US Nationwide distribution.

Code information

All lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0076-2022
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0076-2022. Report a data issue: hello@regdatalab.com.