Z-0087-2023 Class II Ongoing
FDA device recall Z-0087-2023 was initiated by Shanghai United Imaging Healthcare Co., Ltd. on May 30, 2022 and is designated Class II. Reason for recall: The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and… The recall status is ongoing. Affected quantity: 32.