Z-0087-2023 Class II Ongoing

Recalled by Shanghai United Imaging Healthcare Co., Ltd. — Shanghai, N/A

FDA device recall Z-0087-2023 was initiated by Shanghai United Imaging Healthcare Co., Ltd. on May 30, 2022 and is designated Class II. Reason for recall: The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and… The recall status is ongoing. Affected quantity: 32.

Recall details

Product Type
Devices
Report Date
Oct 19, 2022
Initiation Date
May 30, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
32

Product description

Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Reason for recall

The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.

Distribution pattern

Distribution to US states of TX, WI, CA, and MI

Code information

UDI-DI/Serial Numbers: 06971576832026/230002, 230003, 230004, 230005, 230006, 230007, 230008, 230009, 230010, 230011, 230012, 230013, 230014, 230015, 230016, 230017, 230019, 230020, 230021, 230022, 230023, 230024, 230025, 230026, 230027, 230028, 230029, 230030, 230031, 230032, 230033, 232001

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0087-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Shanghai United Imaging Healthcare Co., Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0087-2023. Report a data issue: hello@regdatalab.com.