Z-0091-2021 Class II Terminated

Recalled by Mizuho OSI — Union City, CA

Recall Details

Product Type
Devices
Report Date
October 21, 2020
Initiation Date
October 14, 2019
Termination Date
September 27, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
174 Arms

Product Description

Levo Arm Label: LEVO ARM REF 7887-050 MIZUHO OSI Refurbished Levo Arm Label: LEVO ARM REF 7887-050R MIZUHO OSI - Product Usage: used with Mizuho OSI spinal surgery tables to provide patient head support and positioning during spinal surgery.

Reason for Recall

Due to the potential for fluid ingress into the Arm which could affect the device's electrical system and lead to a failure of the locking mechanism

Distribution Pattern

Worldwide distribution - U.S. Nationwide distribution including in states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, SC, TN, TX, VA, WI, and WV. The countries of Australia, Belgium, Canada, Columbia, Germany, Israel, South Korea, New Zealand, Poland, United Kingdom, and South Africa.

Code Information

Model Number: 7887-050 UDI/GTIN 00842230104992 Affected Serial Numbers: 201 through 452 Model 7887-050R UDI/GTIN 00824230108440 Affected Serial Numbers: 234, 235, 248, 255, 341, and 448