Z-0091-2022 Class I Ongoing

Recalled by Datascope Corp. — Mahwah, NJ

Recall Details

Product Type
Devices
Report Date
October 27, 2021
Initiation Date
September 21, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
131 batteries, 11,517 pumps

Product Description

Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) with the below Model Numbers. 1. Cardiosave Hybrid. Model Numbers: 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-UC-0800-55. 2. Cardiosave Rescue. Model Number: 0998-00-0800-83.

Reason for Recall

A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to customers.

Distribution Pattern

Worldwide distribution. US nationwide, Australia, Brazil, Bahrain, Belgium, Canada, China, Czech Republic, Egypt, France, Germany, India, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, Pakistan, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, UK, United Arab Emirates, Philippines, and Russia

Code Information

Battery Pack Serial Numbers: 192224349IP, 191908720IP, 202369811IP, 202791944IP, 191848115IP, 192099841IP, 192103741IP, 192104141IP, 171218521PE, 181642339PE, 202321308IP, 191912520IP, 191874716IP, 191922620IP, 181642639PE, 192185446IP, 171277935PE, 192190247IP, 202400219IP, 202323808IP, 191910720IP, 181661142PE, 192245250IP, 202599332IP, 181645841PE, 181526530PE, 202403719IP, 181705244PE, 192104541IP, 191787312IP, 181646041PE, 191955422IP, 202389919IP, 191822513IP, 191853415IP, 192004724IP, 192196447IP, 192155645IP, 202330609IP, 171325749PE, 192018828IP, 191778911IP, 192187947IP, 192196247IP, 181609237PE, 192098341IP, 191972823IP, 191860615IP, 202719740IP, 202334109IP, 202651035IP, 191852615IP, 192173146IP, 202386519IP, 202398619IP, 202416720IP, 202599432IP, 191956322IP, 192205847IP, 192220349IP, 192081138IP, 202617233IP, 181659242PE, 202266705IP, 171255732PE, 192179046IP, 192179246IP, 202280605IP, 181673143PE, 192083838IP, 192106442IP, 192216649IP, 191956122IP, 181721730IP, 191943120IP, 191971023IP, 181405120PE, 181474326PE, 181440222PE, 192225649IP, 191969623IP, 192220749IP, 181706344PE, 181694644PE, 191968223IP, 202593632IP, 191863015IP, 171264133PE, 191872616IP, 181487828PE, 171337749PE, 181477426PE, 181748032IP, 191790412IP, 191902319IP, 192195547IP, 192200947IP, 191968723IP, 202552831IP, 191806513IP, 202401419IP, 191904319IP, 181492328PE, 202388319IP, 202482724IP, 192241550IP, 171339849PE, 191831214IP, 192084038IP, 181694244PE, 192210249IP, 202725540IP, 171345550PE, 171348550PE, 191915920IP, 202371511IP, 202723440IP, 202367411IP, 202279305IP, 181513630PE, 181703744PE, 202389819IP, 192175646IP, 192107842IP, 202394719IP, 202401619IP, 192204947IP, 202654636IP, 192196647IP, 192200647IP, 202782744IP. 1. Cardiosave Hybrid. Model Number (UDI Number): 0998-00-0800-31 (10607567109053), 0998-00-0800-32 (10607567111117), 0998-00-0800-33 (10607567109008), 0998-00-0800-35 (10607567109107), 0998-00-0800-36 (10607567114187), 0998-00-0800-45 (10607567108421), 0998-00-0800-52 (10607567108438), 0998-00-0800-53 (10607567108391), 0998-00-0800-55 (10607567108414), 0998-00-0800-65 (10607567113432), 0998-UC-0800-31 (10607567109053), 0998-UC-0800-33 (10607567109008), 0998-UC-0800-52 (10607567108438), 0998-UC-0800-53 (10607567108391), 0998-UC-0800-55 (10607567108414). 2. Cardiosave Rescue. Model Number: 0998-00-0800-83 (10607567108407).