Z-0092-2014 Class II Terminated

Recalled by Stryker Instruments Div. Of Stryker Corporation — Portage, MI

FDA device recall Z-0092-2014 was initiated by Stryker Instruments Div. Of Stryker Corporation on Oct 2, 2013 and is designated Class II. Reason for recall: It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing possess. The potting material is used to insulate, stabilize and provide a moisture barrier… The recall status is terminated (terminated Apr 15, 2014). Affected quantity: 67.

Recall details

Product Type
Devices
Report Date
Nov 6, 2013
Initiation Date
Oct 2, 2013
Termination Date
Apr 15, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
67

Product description

REF 5100-088, TPS Micro Driver, For Use with Stryker Total Performance System Caution Federal Law (USA) restricts the device to use by or on the order of a physician Driver, Wire and Bone Drill; The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental, orthopedic, maxillofacial, spinal, and plastic surgery. The instruments are also used in the placement of screws, wires, pins, and other fixation devices.

Reason for recall

It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing possess. The potting material is used to insulate, stabilize and provide a moisture barrier to the TPS MicroDriver main Printed Circuit Board Assembly (PCBA). Inadequate cure of the silicone potting could potentially allow it to leak to the outside of the handpiece or cause an electrical ma

Distribution pattern

Worldwide distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Italy, Japan, Spain, Switzerland, and United Kingdom.

Code information

10933, 11433, 11483, 20553, 30163, 31553, 40713, 40833, 41113, 50463, 60133, 60423, 90383, 110093, 110193, 1030133, 1030403, 1040413, 2010013, 2010123, 96090403, 96091343, 97020193, 97030013, 97030543, 97050243, 97060163, 97080033, 98010083, 98020703, 98030103, 98040213, 98060023, 98061033, 98070603, 98090023, 98090473, 98090503, 98101143, 98110323, 99010433, 99030253, 99050033, 99050383, 99060163, 99060423, 99061063, 99061283, 99070453, 99071293, 99080243, 99080483, 99090913, 99101273, 221303733, 326102033, 434909453, 500800103, 504900233, 631112793, 631414853, 700314923, 723606673, 725315933, 735502183, 813304543, and 921700523 ,

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0092-2014
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0092-2014. Report a data issue: hello@regdatalab.com.