Z-0093-2018 Class II Terminated
Recalled by Siemens Medical Solutions USA, Inc.
— Hoffman Estates, IL
FDA device recall Z-0093-2018 was initiated by Siemens Medical Solutions USA, Inc. on Nov 7, 2017 and is designated Class II. Reason for recall: The Broad Quantification option of the Symbia product software version VB20A may not allow modification of two data input values. The failure occurs when the system with the Broad Quantification… The recall status is terminated (terminated Apr 29, 2019). Affected quantity: 2 units.
Recall details
- Product Type
- Devices
- Report Date
- Nov 22, 2017
- Initiation Date
- Nov 7, 2017
- Termination Date
- Apr 29, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 units
Product description
Symbia Intevo 16, SPECT/CT System, Emission Computed Tomography System To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. CT: The CT component is intended to produce cross sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial place taken at different angles or spiral planes taken at different angles.SPECT+CT: Perform CT scans and nuclear imaging studies within the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy. These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
Reason for recall
The Broad Quantification option of the Symbia product software version VB20A may not allow modification of two data input values. The failure occurs when the system with the Broad Quantification option is calibrated with phantoms that do not have a volume of 6500ml and/or when residual dose needs to be entered.
Distribution pattern
Worldwide Distribution to Malaysia, France, Japan, Reunion Island
Code information
Serial Number: 2136, 2142
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0093-2018
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Siemens Medical Solutions USA, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Siemens Medical Solutions USA, Inc. →
Use it in ChatGPT
Get a free API key
How the connector works