Z-0099-2024 Class II Ongoing
FDA device recall Z-0099-2024 was initiated by Alcon Research, LLC on Aug 18, 2023 and is designated Class II. Reason for recall: Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure… The recall status is ongoing. Affected quantity: 3.
Recall details
- Product Type
- Devices
- Report Date
- Oct 18, 2023
- Initiation Date
- Aug 18, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3
Product description
Constellation Vision System, REF: 8065751150
Reason for recall
Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.
Distribution pattern
US: GA, IL, PR
Code information
UDI-DI: 00380657511501, Serial Numbers: 1403028701X, 1202865201X, 0802983301X
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0099-2024
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
Need the full FDA record for Alcon Research, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.