Z-0099-2024 Class II Ongoing

Recalled by Alcon Research, LLC — Irvine, CA

FDA device recall Z-0099-2024 was initiated by Alcon Research, LLC on Aug 18, 2023 and is designated Class II. Reason for recall: Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure… The recall status is ongoing. Affected quantity: 3.

Recall details

Product Type
Devices
Report Date
Oct 18, 2023
Initiation Date
Aug 18, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product description

Constellation Vision System, REF: 8065751150

Reason for recall

Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.

Distribution pattern

US: GA, IL, PR

Code information

UDI-DI: 00380657511501, Serial Numbers: 1403028701X, 1202865201X, 0802983301X

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0099-2024
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Alcon Research, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works