Z-0100-2017 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-0100-2017 was initiated by Roche Diagnostics Operations, Inc. on Jul 15, 2016 and is designated Class II. Reason for recall: Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to… The recall status is terminated (terminated Jun 5, 2017). Affected quantity: 9014 in total.

Recall Details

Product Type
Devices
Report Date
Nov 2, 2016
Initiation Date
Jul 15, 2016
Termination Date
Jun 5, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9014 in total

Product Description

Roche COBAS INTEGRA 800 /800 CTS Analyzer Analyzer, Chemistry (Photometric, Discrete for Clinical use Product Usage: The system cobas p612 is a computer-controlled stand-alone system for aliquoting and sorting of barcoded sample tubes. The system operates with an online connection to the host.

Reason for Recall

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

Distribution Pattern

US Nationwide Distribution

Code Information

Not applicable

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0100-2017
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.