Z-0101-2015 Class II Terminated

Recalled by Estill Medical Technologies, Inc — Dallas, TX

FDA device recall Z-0101-2015 was initiated by Estill Medical Technologies, Inc on Jul 25, 2014 and is designated Class II. Reason for recall: Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in… The recall status is terminated (terminated Aug 21, 2015). Affected quantity: 6,886.

Recall details

Product Type
Devices
Report Date
Oct 29, 2014
Initiation Date
Jul 25, 2014
Termination Date
Aug 21, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6,886

Product description

Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the Thermal Angel Blood and IV Fluid Infusion Warmer. The TA-UB1 battery was designed to provide sufficient power for the Thermal Angel to warm and deliver one liter of blood or IV fluid infusion to a trauma patient.

Reason for recall

Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t

Distribution pattern

Worldwide Distribution - USA (nationwide) and Internationally to Serbia.

Code information

06-2011; 07-2011; 08-2011; 09-2011; 03-2012; 04-2012; 05-2012; 06-2012; 07-2012; 08-2012; 01-2013; 03-2013; 08-2013; 09-2013; 10-2013; 11-2013; 12-2013; 01-2014; 02-2014; 04-2014; 07-2012; 02-2014

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0101-2015
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Estill Medical Technologies, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Estill Medical Technologies, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Estill Medical Technologies, Inc →

Source: openFDA record for Z-0101-2015. Report a data issue: hello@regdatalab.com.