Z-0101-2019 Class II Terminated

Recalled by Agfa Healthcare NV — Mortsel

FDA device recall Z-0101-2019 was initiated by Agfa Healthcare NV on Aug 3, 2018 and is designated Class II. Reason for recall: Adding an additional comparison study to the Clinical Side Bar in the Enterprise Imaging Diagnostics Desktop, can result in a change of the sort order of the studies in the Clinical Side Bar. The… The recall status is terminated (terminated Apr 3, 2020). Affected quantity: 17.

Recall details

Product Type
Devices
Report Date
Oct 17, 2018
Initiation Date
Aug 3, 2018
Termination Date
Apr 3, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
17

Product description

Enterprise Imaging Desktops Enterprise Imaging Desktops (Radiology Suite/Cardiology Suite) when deployed as a departmental Picture Archiving and Communications System (PACS) provides an interface for the acquisition, display, digital processing, annotation, review, printing, storage and distribution of multimodality medical images, reports and demographic information for diagnostic purposes within the system and across computer networks.

Reason for recall

Adding an additional comparison study to the Clinical Side Bar in the Enterprise Imaging Diagnostics Desktop, can result in a change of the sort order of the studies in the Clinical Side Bar. The thumbnail images of the studies are no longer chronologically sorted with the most recent study at the top.

Distribution pattern

US distribution to states of: CA, FL, IN, LA, MN, PA, SC, TX, VA, and Puerto Rico and internationally to: Canada.

Code information

Versions 8.0.0, 8.0.0 SP1 & 8.0.0 SP2., 8.0.1, 8.0.1 SP1, 8.0.1 SP2, 8.0.1 SP3, 8.0.1 SP4, 8.0.1 SP5, 8.0.1 SP6., 8.1, 8.1 SP1, 8.1 SP2.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0101-2019
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0101-2019. Report a data issue: hello@regdatalab.com.