Z-0105-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 21, 2020
- Initiation Date
- August 31, 2020
- Termination Date
- July 1, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 368
Product Description
Percutaneous Neonatal Pigtail Nephrostomy Set, Reference Part Number 080106, Order Number G14508. The set contains a 22 gage, 5 cm Chiba Needle; an 18 gage, 5 cm trocar needle; an .038-inch (.97 mm) diameter, 30 cm PTFE-coated stainless steel; a 5, 6, and 7 French, 10 cm radiopaque dilator; a 6 French, 7 cm radiopaque pigtail catheter; a retention disc with pull tie; and a 10 French, 10 cm clear, polyvinylchloride connecting tube with one-way stopcock.
Reason for Recall
The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation.
Distribution Pattern
Worldwide distribution - US Nationwide.
Code Information
All lots