Z-0107-2027 Class II Ongoing

Recalled by POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC — Cortlandt Manor, NY

FDA device recall Z-0107-2027 was initiated by POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC on Aug 25, 2026 and is designated Class II. Reason for recall: During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected… The recall status is ongoing. Affected quantity: 448400.

Recall details

Product Type
Devices
Report Date
Oct 14, 2026
Initiation Date
Aug 25, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
448400

Product description

SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000;

Reason for recall

During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).

Distribution pattern

Worldwide - US Nationwide distribution in the states and territory of AK, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV and the country of United Kingdom.

Code information

Model/Catalog No. S-100 and S-1000; UDI-DI: N/A; Lot Number E1150A; Expiration Date 06/30/27

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0107-2027
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0107-2027. Report a data issue: hello@regdatalab.com.