Z-0108-2021 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-0108-2021 was initiated by Siemens Medical Solutions USA, Inc on Aug 3, 2020 and is designated Class II. Reason for recall: Potential exists for the collision supervision not to work properly following loss of the individual room configuration data settings causing danger of collision with fix mounted room equipment, walls… The recall status is terminated (terminated Feb 3, 2021). Affected quantity: 124 U.S.

Recall details

Product Type
Devices
Report Date
Oct 21, 2020
Initiation Date
Aug 3, 2020
Termination Date
Feb 3, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
124 U.S

Product description

Artis zee or Artis Q systems (listed below) - Product Usage: for single and biplane diagnostic imaging and interventional procedures. System Name/Material Number: Artis Q BIPLANE 10848282 ARTIS Q CEILING 10848281 Artis Q floor 10848280 ARTIS Q ZEEGO 10848283 ARTIS Q.ZEN BIPLANE 10848355 ARTIS Q.zen ceiling 10848354 Artis Q.zen floor 10848353 Artis zee Biplane 10094141 Artis zee ceiling 10094137 Artis zee floor 10094135 Artis zee floor 10094135 Artis zee multi-purpose 10094139 Artis zeego 10280959

Reason for recall

Potential exists for the collision supervision not to work properly following loss of the individual room configuration data settings causing danger of collision with fix mounted room equipment, walls or floor may occur. This might cause components of the system to fall or tilt which could result in crushing of patients, operators or staff, collision of system parts with patients, as well as result in delay or interruption of the clinical procedure

Distribution pattern

US Natonwide distribution.

Code information

Software VD11C Patch distributed earlier via Update Instructions AX030/19/S or AX051/19/P. Serial Numbers: 103222 103285 103286 105016 109016 109021 109069 109105 109107 109158 109163 109189 109221 109300 109331 109367 109371 109393 109416 109451 109460 109504 109505 109520 111005 111104 111106 111117 111120 111123 111125 117014 117030 117117 117123 117129 117134 117303 117309 117310 117314 117325 117329 117330 121008 121058 121062 121071 121375 121377 121378 121383 121405 121417 121442 123010 123014 123023 123049 123208 123209 123233 124286 125201 125233 137245 137421 137438 137479 137729 138102 138104 138147 138156 138201 146323 147516 147612 147652 147670 147676 147724 147745 147880 147905 147923 147925 147926 147928 147944 148207 148282 148349 154532 154612 154633 154663 154739 155151 157704 158032 158458 158474 158552 160431 160449 160494 160509 160845 160850 160861 160885 160913 160942 160953 160959 160978 160999 161004 161008 161021 161050 161224 161226

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0108-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Medical Solutions USA, Inc →

Source: openFDA record for Z-0108-2021. Report a data issue: hello@regdatalab.com.