Z-0108-2026 Class II Ongoing
Recalled by MICROVENTION INC.
— Aliso Viejo, CA
FDA device recall Z-0108-2026 was initiated by MICROVENTION INC. on Sep 5, 2025 and is designated Class II. Reason for recall: Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel… The recall status is ongoing. Affected quantity: 467.
Recall details
- Product Type
- Devices
- Report Date
- Oct 22, 2025
- Initiation Date
- Sep 5, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 467
Product description
Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED3511, FRED5014, FRED3516, MV-F451827; FRED X 27, REF: XFRED4017-MVE, XFRED4528-MVE, MV-F501427X, XFRED4518-MVE, XFRED5526-MVE, XFRED4539-MVE, FREDX5522-PMA, FREDX3522-PMA, FREDX4525-PMA, FREDX5015-PMA, FREDX5532-PMA, FREDX4013-PMA, FREDX3513-PMA, FREDX4515-PMA, FREDX3517-PMA, MV-F352427X
Reason for recall
Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences.
Distribution pattern
Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL, CO, NE, NY and the countries of DE, ES, PL, GB, IT, IN, AU, KZ, JP, AT, SI, RS, GR, TR, AR, CH, SG, KR, SK, MX, DK, IR, SA, BY, PA, VN, ZA, BR, TW, CN, BE, FR, VE, MY.
Code information
REF/UDI-DI(Lot): FRED5019/00811425020388(0000235839), 00842429117163(0000544498); FRED5029/00811425020395(0000256343), 00842429117170(0000507209); FRED4528/00811425020425(0000265666); FRED4026/00811425021033(0000333125, 0000576973); FRED5526/00842429117194(0000361117); FRED4508/00842429117095(0000372088); FRED5514/00842429117187(0000429786, 0000478385); FRED3536/00842429117033(0000460268, 0000445555); FRED3511/00842429117002(0000452923); FRED5014/00842429117156(0000563591, 0000464401); FRED3516/00842429117019(0000513146, 0000531733, 0000536217); MV-F451827/04987892121906(0000565094); XFRED4017-MVE/00842429107508(0000478838); XFRED4528-MVE/00842429107560(0000518323, 0000663751); MV-F501427X/04987892151231(0000579031); XFRED4518-MVE/00842429107553(0000663074, 0000663750); XFRED5526-MVE/00842429107638(0000704203); XFRED4539-MVE/00842429107577(00007311620); FREDX5522-PMA/00842429114483(0000265641, 0000469609, 0000731155); FREDX3522-PMA/00842429114315(0000469598); FREDX4525-PMA/00842429114414(0000469603); FREDX5015-PMA/00842429114445(0000469606); FREDX5532-PMA/00842429114490(0000469610); FREDX4013-PMA/00842429114346(0000469613, 0000663758, 0000668189); FREDX3513-PMA/00842429114292(0000478834); FREDX4515-PMA/00842429114391(0000478840); FREDX3517-PMA/00842429114308(0000731304); MV-F352427X/04987892151101(0000760963)
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0108-2026
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for MICROVENTION INC.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works