Z-0109-2027 Class III Ongoing
FDA device recall Z-0109-2027 was initiated by Mindray DS USA, Inc. on Aug 31, 2026 and is designated Class III. Reason for recall: Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions The recall status is ongoing. Affected quantity: 32,184 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 14, 2026
- Initiation Date
- Aug 31, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 32,184 units
Product description
ePM Series Patient Monitor. Patient Monitor Series: ePM 10M/ePM 12M/ePM 12MA/ ePM 15M/ePM 10A
Reason for recall
Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Distribution pattern
Worldwide - Nationwide and the country of Canada.
Code information
Part numbers of ePM Series Patient Monitors Operator's Manuals: 1. Guardian Plus Operator's Manual (OEM, MDPro) up to and including revision 4.0; Part No. 046-026408-00 2. ePM 10A Operator's Manual up to and including revision 12.0; Part No. 046-019796-00 3. ePM 12MA Operator's Manual up to and including revision 13.0; Part No. 046-019797-00 4. ePM 10M, ePM 12M Operator's Manual up to and including revision 13.0; Part No. 046-019798-00 5. ePM 15M Operator's Manual up to and including revision 3.0; Part No. 046-028915-00
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0109-2027
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026