Z-0111-2016 Class II Terminated
Recalled by DePuy Orthopaedics, Inc.
— Warsaw, IN
FDA device recall Z-0111-2016 was initiated by DePuy Orthopaedics, Inc. on Sep 8, 2015 and is designated Class II. Reason for recall: DePuy Orthopaedics, Inc. is initiating a voluntary recall of select lots of the Specialist 2 Intramedullary Rod 400mm Instrument (PN 96-6120) due to complaints of the rod fracturing and leaving a… The recall status is terminated (terminated Jul 3, 2018). Affected quantity: 7488.
Recall Details
- Product Type
- Devices
- Report Date
- Oct 14, 2015
- Initiation Date
- Sep 8, 2015
- Termination Date
- Jul 3, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7488
Product Description
Specialist 2 Intramedullary (SP2 IM) Rod 400mm Instrument (PN 96-6120) The SP2 IM Rod 400mm Instrument (PN 96-6120) is used in both primary and revision P.F.C.¿ SIGMA¿ knee procedures to align the femoral locating device and distal femoral cutting block. It is also used with the IM tibial resection. This rod is included with the P.F.C.¿ SIGMA¿ Specialist¿ 2 (SP2 / SPII) Primary Knee Instruments, the P.F.C.¿ SIGMA¿ Specialist ¿ 2 (SP2 / SPII) Revision Knee Instruments, and P.F.C. ¿ SIGMA¿ High Performance (HP) Primary Knee Instruments
Reason for Recall
DePuy Orthopaedics, Inc. is initiating a voluntary recall of select lots of the Specialist 2 Intramedullary Rod 400mm Instrument (PN 96-6120) due to complaints of the rod fracturing and leaving a portion of the rod in patients.
Distribution Pattern
Worldwide Distribution - US including AL; AZ; CA; CO; CT; DC; FL; GA; HI; IA; IL; IN; KY; LA; MA; MD; ME; MI; MN; MO; MS; MT; NC; NE; NH; NJ; NV; NY; OH; OR; PA; RI; SC; TN; TX; UT; VA; VT; WA; WI; and WV and Internationally to ARGENTINA; AUSTRALIA; AUSTRIA; BELGIUM; BRAZIL; CANADA; CHILE; COLOMBIA; CZECH REPUBLIC; EGYPT; FRANCE; GERMANY; GREECE; INDIA; ISRAEL; ITALY; KOREA; Mexico; NEW ZEALAND; NORWAY; P R CHINA; PANAMA; PERU; POLAND; RUSSIA; SINGAPORE; SLOVENIA; SOUTH AFRICA; SWITZERLAND; UAE; UK; URUGUAY; VENEZUELA; and VIETNAM.
Code Information
SP2 IM Rod 400mm Instruments (96-6120) manufactured with 455 stainless steel (SS) or 17-4 SS; CATALOG NO. : 96-6120; GTIN NO.: 10603295246893; Lots: Manufactured with 455 Stainless Steel, Labeled Lot #s-C98CS4; DE5P34; DE5RP4; D95AN4; EB5FV4; FA4G94; DE1FA4; DF4H44; DG9LK4; EB5GH4; EC9JY4; EF4DJ4; FD8MP4; FH8JA4; DG9L64; DJ5E34; DK3E34; EJ7AP4; DK3FE4; EJ7A34; ES2G64; ES2HA4; FH8JX4; ES2HY4; EX5L44; C3JHN4; EX5MS4; C3JHS4; E2SD44; C3JHF4; C4GA54; C4GBT4; C4GCC4; C52F14; C52F74; C52GL4; C52GV4; C67N14; C98BF4; TBACC; TBACZ; FJ4E74; TBCOJ; & Manufactured with 17-4 Stainless Steel- Etched Lot #s- H0797; H0100; H0199; H0200; H0298; H0301; H0396; H0399; H0401; H0495; H0499; H0530; H0596; H0598; H0600; H0601; H0696; H0699; H0796; H0797; H0800; H0899; H0900; H0997; H0998; H1095; H1097; H1195; H1198; H1297; H1298; H1299
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0111-2016
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab
- Oct 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.