Z-0115-2021 Class II Terminated
Recalled by Siemens Healthcare Diagnostics, Inc
— East Walpole, MA
FDA device recall Z-0115-2021 was initiated by Siemens Healthcare Diagnostics, Inc on Aug 25, 2020 and is designated Class II. Reason for recall: Siemens confirmed that the Upper Limit of Normal (ULN) as claimed in the Instructions for Use (IFU) was no longer achieved. This can potentially result in false positive or false negative results. The recall status is terminated (terminated Mar 10, 2023). Affected quantity: Atellica IM H2n Assay =135 kits; Atellica IM H2n Calibrator = 46 kits; ADVIA Centaur Assay = 1,823 kits; ADVIA Centaur…
Recall details
- Product Type
- Devices
- Report Date
- Oct 28, 2020
- Initiation Date
- Aug 25, 2020
- Termination Date
- Mar 10, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- Atellica IM H2n Assay =135 kits; Atellica IM H2n Calibrator = 46 kits; ADVIA Centaur Assay = 1,823 kits; ADVIA Centaur calibrator = 161 kits
Product description
Atellica IM HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10995591, UDI Codes: (01)00630414599151(10)98034139(17)20210130 and (01)00630414599151(10)32478139(17)20210130 Atellica IM HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10995592, UDI Code (01)00630414599168(10)22695A49(17)20201211 ADVIA Centaur HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10308994, UDI codes : (01)00630414471235(10)43157138(17)20210130 (01)00630414471235(10)81653138(17)20210130 (01)00630414471235(10)99728138(17)20210130 (01)00630414471235(10)24043138(17)20210130 (01)00630414471235(10)32477138(17)20210130 Advia Centaur Systems HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10308993, UDI Codes: (01)00630414471228(10)64661A48(17)20201112 (01)00630414471228(10)81211A48(17)20201112 (01)00630414471228(10)91743A48(17)20201112 (01)00630414471228(10)22124A49(17)20201211 Product Usage: This product is intended to be used for in vitro diagnostic use in calibrating the ADVIA Centaur systems HER-2/neu assay.
Reason for recall
Siemens confirmed that the Upper Limit of Normal (ULN) as claimed in the Instructions for Use (IFU) was no longer achieved. This can potentially result in false positive or false negative results.
Distribution pattern
US Nationwide distribution.
Code information
Atellica IM HER-2/neu Assay 50 Test Kit: Siemens Material Number (SMN) 10995591; EXP Date 2021-01-30; Lot # 98034139 w/ UDI (01)00630414599151(10)98034139(17)20210130; Lot # 32478139 w/ UDI (01)00630414599151(10)32478139(17)20210130 Atellica IM HER-2/neu Calibrator (2 pack): SMN 10995592; EXP Date 2020-12-11; Lot # 22695A49; UDI (01)00630414599168(10)22695A49(17)20201211 ADVIA Centaur HER-2 Assay 50 Test Kit: SMN 10308994; EXP Date 2021/01/30; Lot # 43157138 w/ UDI (01)00630414471235(10)43157138(17)20210130 ; Lot# 81653138 w/ UDI (01)00630414471235(10)81653138(17)20210130 ; Lot# 99728138 w/ UDI (01)00630414471235(10)99728138(17)20210130 ; Lot # 24043138 w/ UDI (01)00630414471235(10)24043138(17)20210130 ; Lot #32477138 w/ UDI (01)00630414471235(10)32477138(17)20210130 ; ADVIA Centaur HER-2/neu Calibrator (2 pack): SMN 10308993; EXP Date 2020-11-12 Lot # 64661A48 w/ UDI (01)0063041447(10)64661A48(17)20201112 Lot # 81211A48 w/ UDI (01)0063041447(10)81211A48(17)20201112 Lot# 91743A48 w/ UDI (01)0063041447(10)91743A48(17)20201112 EXP Date 2020-12-11; Lot # 22124A49 w/ UDI (01)0063041447(10)22124a49(17)20201112
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0115-2021
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.