Z-0117-2013 Class II Terminated
Recalled by Advanced Bionics Corporation
— Sylmar, CA
FDA device recall Z-0117-2013 was initiated by Advanced Bionics Corporation on Sep 14, 2007 and is designated Class II. Reason for recall: Advanced Bionics is recalling the the HiRes 90K cochlear implant with HiFocus Helix Electrode because of loose-fitting insertion tool assemblies. The recall status is terminated (terminated Oct 22, 2012). Affected quantity: 17 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 31, 2012
- Initiation Date
- Sep 14, 2007
- Termination Date
- Oct 22, 2012
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 17 units
Product description
HiRes 90K devices with Helix Electrodes Model CI 1400-02H The HiResolution Bionic Ear System is a cochlear implant system designed to provide useful hearing to individuals with severe to profound hearing loss. It consists of internal and external components. The internal components include a stimulator (HiRes90K ICS) and electrode array (HiFocus 1j or Helix). The HiRes90K ICS is implanted surgically under the skin behind the ear. The electrode array is inserted into the cochlea. The external components include a sound processor (body-worn or ear-level), a headpiece, and a cable. The system converts sound into electrical energy that activates the auditory nerve. The auditory nerve then sends information to the brain, where it is interpreted as sound. The Helix consists of a fantail, electrode lead, and Helix electrode array. The electrode, comprised of platinum-iridium wires, is housed in a silicone carrier and extends from the titanium case. The Helix electrode array is 24.5 mm in length and is designed to be inserted approximately 18-21 mm into a normal patent cochlea. It consists of 16 planar contacts arranged along the medial surface of the electrode array for stimulation of discrete segments of the cochlea. The Helix electrode insertion tool is designed to assist the surgeon with the insertion of the Helix electrode. Only the Helix insertion should be used for insertion of the Helix electrode; the use of other instruments may result in damage to the electrode. The insertion tool is a single-use instrument.
Reason for recall
Advanced Bionics is recalling the the HiRes 90K cochlear implant with HiFocus Helix Electrode because of loose-fitting insertion tool assemblies.
Distribution pattern
Worldwide distribution: USA (nationwide) including state: IL and country of: Europe.
Code information
Serial Numbers: 351323, 251153, 350989, 351226, 350196, 351136, 350140, 350201, 351261, 351264, 351266, 351269, 351412, 351415, 351429, 350157, 351303.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0117-2013
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Advanced Bionics Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Advanced Bionics Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Advanced Bionics Corporation →
Use it in ChatGPT
Get a free API key
How the connector works