Z-0119-2015 Class II Terminated
Recalled by Aerogen Ltd.
— Galway
FDA device recall Z-0119-2015 was initiated by Aerogen Ltd. on September 2, 2014 and is designated Class II. Reason for recall: Aerogen has received a small number of complaints in relation to the main power adapter manufactured by FRIWO GmbH and used in the Aeroneb Solo and Pro Nebuliser systems. The complaints received ident… The recall status is terminated (terminated April 25, 2016). Affected quantity: 1,223 power adapters.
Recall Details
- Product Type
- Devices
- Report Date
- October 29, 2014
- Initiation Date
- September 2, 2014
- Termination Date
- April 25, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,223 power adapters
Product Description
The Aeroneb Professional Nebulizer System is a portable medical device for multiple patient use that is intended to aerosolize physician-prescribed solutions for inhalation to patients on and off ventilation or other positive pressure breathing assistance. The Aeroneb Professional Nebulizer System is suitable for use in adult and pediatric patients. The Aeroneb Solo Nebulizer System is an iteration of the Aeroneb Professional Nebulizer System which is single patient use. The FRIWO AC/DC adapter is a component of the Aerogen Nebulizer systems which is used to power the Pro and Pro X controllers.
Reason for Recall
Aerogen has received a small number of complaints in relation to the main power adapter manufactured by FRIWO GmbH and used in the Aeroneb Solo and Pro Nebuliser systems. The complaints received identified that the power adapter cover had broken and dislodged from the body of the adapter prior to use.
Distribution Pattern
Worldwide Distribution-USA (nationwide) including the states of CA, GA, IL, MS, NC, and TX, and the countries of Australia, Belgium, Canada, China, Czech Republic, Denmark, Egypt, Finland, France, Germany, Hungary, India, Ireland, Israel, Italy, Kuwait, Libya, Malaysia, Netherlands, Norway, Poland, Qatar, Russia, South Korea, Singapore, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Code Information
FRIWO AC/DC Adapter Model Number: GPP10; The FRIWO AC/DC Adapter is supplied with the following Aeroneb Nebulizer systems and their respective model numbers: 1) Model Number: AG-AS3000-AU; Lot Numbers: 9144140519057 , 9144140616062 , 9144140702155 , 9144140710034; 2) Model Number: AG-AS3000-IN; Lot Numbers: 9125140522116 , 9125140526101 , 9125140527077 , 9125140609002 , 9125140609007 , 9125140616064 , 9125140627057 , 9125140702166 , 9125140710031 , 9125140805010 , 9125140805134; 3) Model Number: AG-AS3000-NE; Lot Numbers: 9122140520089 , 9122140521089 , 9122140528119 , 9122140609041 , 9122140626100 , 9122140728097 , 9122140730088; 4) Model Number: AG-AS3000-SC; Lot Numbers: 9124140609011 , 9124140620092 , 9124140623091; 5) Model Number: AG-AS3000-SE; Lot Numbers: 9123140521074 , 9123140610023 , 9123140710033 , 9123140724097; 6) Model Number: AG-AS3000-UK; Lot Numbers: 9120140519028 , 9120140519061 , 9120140526111 , 9120140630135 , 9120140701144 , 9120140702156 , 9120140714044; 7) Model Number: AG-AS3000-US; Lot Numbers: 9121140521090 , 9121140522091 , 9121140522093 , 9121140523094 , 9121140613066 , 9121140630138 , 9121140707023 , 9121140709030; 8) Model Number: AG-PX1000-IN; Lot Numbers: 9133140704002; 9) Model Number: ; Lot Numbers: ; 10) Model Number: AG-PX1000-NE; Lot Numbers: 9130140528119 , 9130140616060 , 9130140730040; 11) Model Number: AG-PX1000-SC; Lot Numbers: 9132140519072 , 9132140526107; 12) Model Number: AG-PX1000-SE; Lot Numbers: 9131140623154; 13) Model Number: AG-PX1000-UK; Lot Numbers: 9128140519081 , 9128140630135 , 9128140709024; 14) Model Number: AG-PX1000-US; Lot Numbers: 9129140527093 , 9129140617070 , 9129140714047 , 9129140721087; 15) Model Number: AG-PX1000-US-Promo; Lot Numbers: 9129140522090 , 9129140522091 , 9129140527093 , 9129140528095 , 9129140528096 , 9129140616066 , 9129140617067 , 9129140617069 , 9129140617070 , 9129140617071