Z-0121-2017 Class II Terminated

Recalled by Synergetics Inc — O Fallon, MO

Recall Details

Product Type
Devices
Report Date
October 26, 2016
Initiation Date
August 11, 2016
Termination Date
April 27, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
255 units

Product Description

25 gauge VersaVit 2.0 Procedure Pack with Photon Wide Field Illuminator, Product/Part Number 70025SP, packaged in HIPS rigid tray with a Tyvek lid sterile barrier system.

Reason for Recall

Vitrectomy cutters have been found not to be operating within the correct pressure pulse per specification.

Distribution Pattern

Devices were distributed to KY, VA, TN, OH, IN, CA, NJ, NY, TX, SC, AL, MS, MI, and UT. Foreign distribution to the UK, France, Mexico, and India.

Code Information

Lot number M543990 (1 each or box of 6 each), exp 10/2018 Lot number M545780 (1 each or box of 6 each), exp 10/2018 Lot number M569020 (1 each or box of 6 each), exp 01/2019 Lot number M578140 (1 each or box of 6 each), exp 02/2019