Z-0124-2022 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics, Inc — East Walpole, MA

Recall Details

Product Type
Devices
Report Date
October 27, 2021
Initiation Date
September 28, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1193

Product Description

Atellica IM 1300 Analyzer, SMN 11066001

Reason for Recall

Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea level. As a result, customers at these altitudes may observe an increased number of Reagent Volume Check Errors.

Distribution Pattern

Domestic distribution to AZ CA CO GA ID KS MO MT NC NE NM NV OH OK OR PA TN TX UT VA WA WI WY. Foreign distribution worldwide.

Code Information

All units