Z-0125-2022 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics, Inc — East Walpole, MA

FDA device recall Z-0125-2022 was initiated by Siemens Healthcare Diagnostics, Inc on Sep 28, 2021 and is designated Class II. Reason for recall: Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea level. As… The recall status is ongoing. Affected quantity: 1733.

Recall details

Product Type
Devices
Report Date
Oct 27, 2021
Initiation Date
Sep 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1733

Product description

Atellica IM 1600 Analyzer, SMN 11066000

Reason for recall

Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea level. As a result, customers at these altitudes may observe an increased number of Reagent Volume Check Errors.

Distribution pattern

Domestic distribution to AZ CA CO GA ID KS MO MT NC NE NM NV OH OK OR PA TN TX UT VA WA WI WY. Foreign distribution worldwide.

Code information

All units.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0125-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Healthcare Diagnostics, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Healthcare Diagnostics, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Healthcare Diagnostics, Inc →

Source: openFDA record for Z-0125-2022. Report a data issue: hello@regdatalab.com.