Z-0127-2014 Class II Terminated

Recalled by Stryker Instruments Div. Of Stryker Corporation — Portage, MI

FDA device recall Z-0127-2014 was initiated by Stryker Instruments Div. Of Stryker Corporation on Nov 4, 2013 and is designated Class II. Reason for recall: Stryker has identified a potential for the auger and/or cannula to fracture when the Dekompressor is used contrary to the warnings in the Instructions For Use (IFU). Specifically, manually bending the… The recall status is terminated (terminated Apr 29, 2014). Affected quantity: 20,385 total distributed since 2004.

Recall details

Product Type
Devices
Report Date
Nov 13, 2013
Initiation Date
Nov 4, 2013
Termination Date
Apr 29, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20,385 total distributed since 2004

Product description

Dekompressor Percutaneous Discectomy Probe. The Dekompressor is a single use disposable discectomy probe that passes through and works in conjunction with an introducer cannula to remove intervertebral disc nucleus pulposus material.

Reason for recall

Stryker has identified a potential for the auger and/or cannula to fracture when the Dekompressor is used contrary to the warnings in the Instructions For Use (IFU). Specifically, manually bending the cannula can weaken the device and may result in a fracture of the auger and/or cannula. This could result in the need for additional medical intervention to remove the broken piece, pain or loss of m

Distribution pattern

Nationwide Distribution including the states of AR, AZ, CA, CO, FL, GA, IL, KS, MS, ND, and NE.

Code information

Part Number(s) 0407-251-000, 0407-260-000, 0407-265-000, 0407-266-000, 0407-280-000, and 0407-281-000.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0127-2014
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0127-2014. Report a data issue: hello@regdatalab.com.