Z-0137-2024 Class II Ongoing
FDA device recall Z-0137-2024 was initiated by Maquet Cardiovascular, LLC on Sep 13, 2023 and is designated Class II. Reason for recall: Housing Mount Jaw of these devices breaking during normal use and may result in injury to the epicardial tissue or vasculature, or a coronary artery or previously placed bypass graft, or could result… The recall status is ongoing. Affected quantity: 2951 units.
Recall details
- Product Type
- Devices
- Report Date
- Nov 1, 2023
- Initiation Date
- Sep 13, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2951 units
Product description
Maquet Acrobat-i Vacuum Stabilizer System Intended for use during performance of cardiac surgical procedures through a sternotomy incision approach. Product Code: OM-10000Z
Reason for recall
Housing Mount Jaw of these devices breaking during normal use and may result in injury to the epicardial tissue or vasculature, or a coronary artery or previously placed bypass graft, or could result in procedural delay and/or conversion.
Distribution pattern
Nationwide Foreign: AT BE DE ES FR GB IT JP KR LU NL PL PT TR
Code information
UDI-DI: 00607567100008 Lot Numbers: 3000188485, 3000193165, 3000193670, 3000195673, 3000198652, 3000199944, 3000201253, 3000204442
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0137-2024
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab