Z-0139-2025 Class II Ongoing

Recalled by SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION — Kyoto, N/A

Recall Details

Product Type
Devices
Report Date
October 30, 2024
Initiation Date
September 6, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13

Product Description

Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image processing equipment, x-ray high voltage generator, x-ray tube. There are several types of each component.

Reason for Recall

If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.

Distribution Pattern

US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.

Code Information

UDI-DI: 04540217058327, 04540217049066, 04540217058143 Serial Numbers: 41EFCD7AC001, 41E58C773001, 41EFCD79C001, 41EFCD78B001, 41EFC73A9001, 41EFCD7D4001, 41E58C76B001, 41EFC73D9001, 41EFCD77C001, 41EFC73A5001, 41EFC73C5001, 41EFCD79A001, 41EFCD7A6001.