Z-0141-2023 Class I Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

Recall Details

Product Type
Devices
Report Date
November 9, 2022
Initiation Date
September 30, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6407 units

Product Description

Hillrom Centrella Hospital Bed with WatchCare Incontinence Management System, Product code/Part numbers: P7900B100193, P7900B100195, P7900B100230, P7900B100236, P7900B100239, P7900B100244, P7900B100256, P7900B100257, P7900B100259, P7900B100264, P7900B100268, P7900B100269, P7900B100292, P7900B100296, P7900B100302, P7900B100303, P7900B100304, P7900B100305, P7900B100312, P7900B100316, P7900B100318, P7900B100335, P7900B100341, P7900B100342, P7900B100354, P7900B100361, P7900B100368, P7900B100372, P7900B100381, P7900B100382, P7900B100384, P7900B100385, P7900B100387, P7900B100388, P7900B100392, P7900B100395, P7900B100396, P7900B100407, P7900B100408, P7900B100438, P7900B100441, P7900B1PL270, P7900B1PLT71, P7900B1SPL05, P7900B200193, P7900B200195, P7900B200230, P7900B200236, P7900B200264, P7900B200268, P7900B200292, P7900B200302, P7900B200318, P7900B200342, P7900B200361, P7900B200372, P7900B200381, P7900B200382, P7900B200392, P7900B200395, P7900B200396, P7900B200408, P7900B200439, P7900B200441, P7900B200443, P7900B200445, P7900B200452, P7900B200453, P7900B200464, P7900B200485, P7900B2SPL12, P7900B300441, P7900B300453, P7900B300518

Reason for Recall

Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).

Distribution Pattern

US

Code Information

UDI/DI 887761985162, all serial numbers