Z-0142-2022 Class II Terminated

Recalled by Canon Medical System, USA, INC. — Tustin, CA

Recall Details

Product Type
Devices
Report Date
October 27, 2021
Initiation Date
November 11, 2020
Termination Date
June 30, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 units

Product Description

Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A

Reason for Recall

A software problem has been identified which could result in the diagnostic imaging system not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the diagnostic imaging system failing resulting in rescanning and reinjection of contrast medium.

Distribution Pattern

US Nationwide distribution in the states of NV, AR MN.

Code Information

Affected serial numbers: 3DA1982001, 3DA2072002, 3DA2072003