Z-0146-2027 Class II Ongoing
FDA device recall Z-0146-2027 was initiated by SI-BONE, Inc. on Aug 21, 2026 and is designated Class II. Reason for recall: The pelvic fixation screws failed to meet acceptance criteria for tube packaging seal integrity testing; therefore, sterility assurance cannot be ensured. The recall status is ongoing. Affected quantity: 100 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 21, 2026
- Initiation Date
- Aug 21, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 100 units
Product description
iFuse TORQ Implant System Model/Catalog Number: 10055T(10x55mm), 10060T(10x60mm), 10065T(10x65mm), 10080T(10x55mm), 11580T(11.5x80mm)
Reason for recall
The pelvic fixation screws failed to meet acceptance criteria for tube packaging seal integrity testing; therefore, sterility assurance cannot be ensured.
Distribution pattern
US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.
Code information
10055T-UDI:00810055520077/Lot:9142871 10060T-UDI:00810055520084/Lot:9137761 10065T-UDI:00810055520091/Lot:9139213 10080T-UDI:00810055520121/Lot:9137781 11580T-UDI:00810055520367/Lot:9142911