Z-0147-2027 Class II Ongoing
FDA device recall Z-0147-2027 was initiated by INTEGRATED MEDCRAFT LLC on Sep 25, 2026 and is designated Class II. Reason for recall: Subrecall of a kitted device from SafeGuard Medical due to internal testing identified an issue affecting the device's sterile pad, potentially causing the sterile pad to degrade over time. The… The recall status is ongoing. Affected quantity: 1033.
Recall details
- Product Type
- Devices
- Report Date
- Oct 21, 2026
- Initiation Date
- Sep 25, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1033
Product description
FORWARD Gunshot Wound Kit; Model: AFT-GSW;
Reason for recall
Subrecall of a kitted device from SafeGuard Medical due to internal testing identified an issue affecting the device's sterile pad, potentially causing the sterile pad to degrade over time. The degradation may cause remnants of the nonwoven pad material to dislodge on the wound or in close proximity to the wound.
Distribution pattern
US Nationwide distribution in the states of GA, VA.
Code information
Model: ATF-GSW; UDI-DI: 00850056748063; Lot Numbers: IMC-00000024, IMC-00000034, IMC-00000053, IMC-00000056, IMC-00000076, IMC-00000081, IMC-00000082, IMC-00000083, IMC-00000084, IMC-00000085, IMC-00000094, IMC-00000148, IMC-00000149, IMC-00000152, IMC-00000153, IMC-00000154, IMC-00000173, IMC-00000174, IMC-00000179, IMC-00000186, IMC-00000187;