Z-0157-2013 Class II Terminated

Recalled by Sorin Group USA, Inc. — Arvada, CO

FDA device recall Z-0157-2013 was initiated by Sorin Group USA, Inc. on Sep 14, 2012 and is designated Class II. Reason for recall: Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient… The recall status is terminated (terminated Apr 23, 2013). Affected quantity: 1,640 units.

Recall details

Product Type
Devices
Report Date
Oct 31, 2012
Initiation Date
Sep 14, 2012
Termination Date
Apr 23, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,640 units

Product description

Gish 3/8 x 3/8" Straight Connector with Luer fitting, Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15. Product Usage: The connectors are indicated for use in connecting flexible tubing used in extracorporeal blood or fluid circuits. The connectors are distributed as sterile stand alones and bulk non-sterile.

Reason for recall

Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury.

Distribution pattern

US Nationwide Distribution in the states of: AL, AR, CA, CO, CT, FL, IL, MI, MN, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA and WI.

Code information

Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0157-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.