Z-0158-2022 Class II Ongoing

Recalled by Medtronic Xomed, Inc. — Jacksonville, FL

Recall Details

Product Type
Devices
Report Date
October 27, 2021
Initiation Date
August 31, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
274,991 total devices

Product Description

Medtronic Sterile single-use blades that are components of the StraightShot M Series Handpiece and Fusion ENT Navigation Software, and the Midas Rex Microsaw and Triton System with the following product numbers: (1) REF 40A2500, Nucleus Removal Angled Blade, 25 cm x 4.0 mm; (2) REF 40R2000, Nucleus Removal Set, 20 cm x 4.0 mm; (3) REF 40R2500, Nucleus Removal Set, 25 cm x 4.0 mm; (4) REF 40S2000, Nucleus Removal Straight Blade, 20 cm x 4.0 mm; (5) REF 40S2500, Nucleus Removal Straight Blade, 25 cm x 4.0 mm; (6) REF 45A2000, Nucleus Removal Angled Blade, 20 cm x 4.5 mm; (7) REF 45A2500, Nucleus Removal Angled Blade, 25 cm x 4.5 mm; (8) REF 45R2000, Nucleus Removal Set, 20 cm x 4.5 mm; (9) REF 45R2500, Nucleus Removal Set, 25 cm x 4.5 mm; (10) REF 45S2000, Nucleus Removal Straight Blade, 20 cm x 4.5 mm; (11) REF 45S2500, Nucleus Removal Straight Blade, 25 cm x 4.5 mm.

Reason for Recall

There was an increase in complaints related to tip breaks and wobble/vibration.

Distribution Pattern

Worldwide distribution - US Nationwide, including PR. There was also government/military distribution and the countries of Australia, Austria, Bahrain, Belgium, Canada, Chile, Croatia, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Republic of Korea, Kuwait, Latvia, Libya, Luxembourg, Macao, Malaysia, Mauritius, Mongolia, Namibia, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Viet Nam.

Code Information

(1) REF 40A2500, GTIN/UPN 00763000033835, Lot/serial number 0220002825; (2) REF 40R2000, GTIN/UPN 00763000033866, Lot/serial number 0218179458; (3) REF 40R2500, GTIN/UPN 00763000033798, Lot/serial number 0218299192; (4) REF 40S2000, GTIN/UPN 00763000033873, Lot/serial number 0218179459; (5) REF 40S2500, GTIN/UPN 00763000043193, Lot/serial number 0220226674; (6) REF 45A2000, GTIN/UPN 00763000033880, Lot/serial number 0220443669; (7) REF 45A2500, GTIN/UPN 00763000043223, Lot/serial number 0218323602; (8) REF 45R2000, GTIN/UPN 00763000033897, Lot/serial numbers 0218284087, 0218292178, 0218973633; (9) REF 45R2500, GTIN/UPN 00763000033811, Lot/serial number 0218297618; (10) REF 45S2000, GTIN/UPN 00763000033903, Lot/serial number 0218299193; (11) REF 45S2500, GTIN/UPN 00763000043216, Lot/serial number 0218297619.