Z-0159-2016 Class II Terminated

Recalled by LifeHealth, LLC — Roseville, MN

Recall Details

Product Type
Devices
Report Date
November 4, 2015
Initiation Date
September 16, 2015
Termination Date
March 21, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4659 boxes of 25. (116475 cartridges)

Product Description

CC cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 039903. The CC Cartridge are single use, disposable, in vitro diagnostics intended for point of care (POC), professional, use with the IRMA TRUpoint¿ Blood Analysis System.

Reason for Recall

May report values outside the published performance specifications for pH. After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discrepancy is greater as shelf life progresses and can be seen on the aqueous controls before the blood.

Distribution Pattern

US: Nationwide (AL, AR, CA, CO, FL, IL, IN, IA, KS, KY, MD, MA,MI, MN, MS, MO, NE, NJ, NY,NC, OH, OK, PA, TN, TX, VA) OUS: ALBANIA, BELGIUM, BOLIVIA, CANADA, CHILE, CHINA, DENMARK. DOMINICAN REPUBLIC, ECUADOR, EGYPT, ENGLAND, ESTONIA, PHILIPPINES, FRANCE, GERMANY, GREECE, ICELAND, INDONESIA, IRELAND, ITALY, JAMAICA, JAPAN, MOROCCO, NETHERLANDS, NORWAY, PERU, PORTUGAL, SOUTH AFRICA, SOUTH KOREA, SPAIN, TAIWAN, TUNISIA, TURKEY, UNITED ARAB EMIRATES, VIETNAM.

Code Information

ASBEZ ASCXE ASEMB ASHWQ ASJMY ASLGI ASMXR ASOQO ASRXC ASTQR ASWZI