Z-0163-2013 Class II Terminated

Recalled by Stryker Spine — Allendale, NJ

FDA device recall Z-0163-2013 was initiated by Stryker Spine on Jul 5, 2012 and is designated Class II. Reason for recall: Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm. The recall status is terminated (terminated Jan 28, 2013). Affected quantity: 6154 units.

Recall details

Product Type
Devices
Report Date
Nov 7, 2012
Initiation Date
Jul 5, 2012
Termination Date
Jan 28, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6154 units

Product description

Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm, Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cestas France +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Ct., Allendale NJ 07401-1677 USA +1-201-750-8000 Drill bits are reusable instruments, sterilized prior to use, used in conjunction with an Aviator Drill Guide to provide a pathway for screw insertion. The aviator Anterior Cervical Plating (ACP) System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation.

Reason for recall

Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.

Distribution pattern

Nationwide distribution.

Code information

E207315 Catalog number 48770610, 48770612, 48770616 All lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0163-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0163-2013. Report a data issue: hello@regdatalab.com.