Z-0163-2013 Class II Terminated
FDA device recall Z-0163-2013 was initiated by Stryker Spine on Jul 5, 2012 and is designated Class II. Reason for recall: Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm. The recall status is terminated (terminated Jan 28, 2013). Affected quantity: 6154 units.
Recall details
- Product Type
- Devices
- Report Date
- Nov 7, 2012
- Initiation Date
- Jul 5, 2012
- Termination Date
- Jan 28, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6154 units
Product description
Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm, Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cestas France +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Ct., Allendale NJ 07401-1677 USA +1-201-750-8000 Drill bits are reusable instruments, sterilized prior to use, used in conjunction with an Aviator Drill Guide to provide a pathway for screw insertion. The aviator Anterior Cervical Plating (ACP) System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation.
Reason for recall
Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.
Distribution pattern
Nationwide distribution.
Code information
E207315 Catalog number 48770610, 48770612, 48770616 All lots