Z-0163-2026 Class I Ongoing

Recalled by IMPRIS Imaging Inc — Chaska, MN

FDA device recall Z-0163-2026 was initiated by IMPRIS Imaging Inc on Sep 30, 2025 and is designated Class I. Reason for recall: There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Nov 5, 2025
Initiation Date
Sep 30, 2025
Voluntary/Mandated
Voluntary: Firm initiated

Product description

IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA

Reason for recall

There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.

Code information

IMRIS Part Number(s): 120823-000, UDI/DI 00857534006745

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0163-2026
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for IMPRIS Imaging Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for IMPRIS Imaging Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about IMPRIS Imaging Inc →

Source: openFDA record for Z-0163-2026. Report a data issue: hello@regdatalab.com.