Z-0164-2015 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc — East Walpole, MA

FDA device recall Z-0164-2015 was initiated by Siemens Healthcare Diagnostics, Inc on Sep 26, 2014 and is designated Class II. Reason for recall: Positive bias for Calibrator E lot kits ending in 38 and 39 with the ADVIA Centaur Systems Cortisol and Progesterone assays compared to the respective master curves. The recall status is terminated (terminated Jan 29, 2016). Affected quantity: Total = 19,352 (13,881 - 2 pack, 4878 - 6 pack, 593 - 6 pack REF).

Recall details

Product Type
Devices
Report Date
Nov 12, 2014
Initiation Date
Sep 26, 2014
Termination Date
Jan 29, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Total = 19,352 (13,881 - 2 pack, 4878 - 6 pack, 593 - 6 pack REF)

Product description

Siemens Calibrator E; SMN 10309079 (2 pack), SMN 10321075 (6 pack), SMN 10335532 (6 pack Ref)

Reason for recall

Positive bias for Calibrator E lot kits ending in 38 and 39 with the ADVIA Centaur Systems Cortisol and Progesterone assays compared to the respective master curves.

Distribution pattern

Nationwide including Puerto Rico. Worldwide: Afghanistan, Armenia, Austria, Bahrain, Belgium, Bulgaria, Burkina Faso, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Georgia, Germany, Great Britain, Greece, Hungary, Iran, Iraq, Ireland, Italy, Jordan, Kazakhstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malta, Morocco, Netherlands, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Unit.Arab Emir., Vatican, Yemen, AR, AU, BD, BR, CA , CH, CL, CN, CO, DO, EC, EG, GT, HK, ID, IE, IL, IN, JP, KR, LY, MM, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VN, ZA

Code information

26284A38, 26900A38, 27991A38, 28192A38, 29571A38, 31162A38, 32488A38, 35243AB39, 35295AB39, 36091AB39, 39464AB39, 40245AB39, 41317AB39, 26285A38, 27348A38, 29169A38, 30710A38, 31261A38, 32687A38, 35259AB39, 35359AB39, 36289AB39, 39031AB39, 41124AB39, 27236A38, 35051A38, 35839AB39, 40141AB39, 41702AB39

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0164-2015
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Healthcare Diagnostics, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Healthcare Diagnostics, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Healthcare Diagnostics, Inc →

Source: openFDA record for Z-0164-2015. Report a data issue: hello@regdatalab.com.