Z-0166-2013 Class II Terminated
FDA device recall Z-0166-2013 was initiated by Sorin Group USA, Inc. on Sep 20, 2012 and is designated Class II. Reason for recall: Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board… The recall status is terminated (terminated Jan 28, 2013). Affected quantity: 12 units (3 units USA and 9 units OUS).
Recall details
- Product Type
- Devices
- Report Date
- Nov 7, 2012
- Initiation Date
- Sep 20, 2012
- Termination Date
- Jan 28, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 12 units (3 units USA and 9 units OUS)
Product description
SCP Pump Control Panel, Product Code: 60-02-15, Serial Numbers: 60S09121, 60S09123-60S09129, 60S09136-60S09139. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping through cardiopulmonary bypass circuit for typical durations of six hours or less.
Reason for recall
Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.
Distribution pattern
Worldwide distribution: USA (nationwide) including states of : MI and NY; and countries of: Belgium, Canada, Japan, India, Russia, United Arab Emirates.
Code information
Product Code: 60-02-15, Serial Numbers: 60S09121, 60S09123-60S09129, 60S09136-60S09139.