Z-0166-2026 Class II Ongoing

Recalled by Philips North America — Cambridge, MA

FDA device recall Z-0166-2026 was initiated by Philips North America on Sep 12, 2025 and is designated Class II. Reason for recall: Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short. The recall status is ongoing. Affected quantity: 29 units.

Recall details

Product Type
Devices
Report Date
Oct 22, 2025
Initiation Date
Sep 12, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
29 units

Product description

Cardiac Workstation 7000; Model Number: 860441;

Reason for recall

Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.

Distribution pattern

US Nationwide distribution in the states of CA, IA, TN, TX, VA.

Code information

Model Number: 860441; UDI-DI: 00884838094833; Serial Numbers: US42520284, US32520248, US32520244, US22520238, US22520221;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0166-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0166-2026. Report a data issue: hello@regdatalab.com.