Z-0167-2013 Class II Terminated

Recalled by Sorin Group USA, Inc. — Arvada, CO

FDA device recall Z-0167-2013 was initiated by Sorin Group USA, Inc. on Sep 20, 2012 and is designated Class II. Reason for recall: Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board… The recall status is terminated (terminated Jan 28, 2013). Affected quantity: 8 units (2 units USA and 6 units OUS).

Recall details

Product Type
Devices
Report Date
Nov 7, 2012
Initiation Date
Sep 20, 2012
Termination Date
Jan 28, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 units (2 units USA and 6 units OUS)

Product description

SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping through cardiopulmonary bypass circuit for typical durations of six hours or less.

Reason for recall

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.

Distribution pattern

Worldwide distribution: USA (nationwide) including states of : MI and NY; and countries of: Belgium, Canada, Japan, India, Russia, United Arab Emirates.

Code information

Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0167-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0167-2013. Report a data issue: hello@regdatalab.com.