Z-0168-2014 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-0168-2014 was initiated by Roche Diagnostics Operations, Inc. on Sep 16, 2013 and is designated Class II. Reason for recall: Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the… The recall status is terminated (terminated May 12, 2015). Affected quantity: 527 units.

Recall Details

Product Type
Devices
Report Date
Nov 13, 2013
Initiation Date
Sep 16, 2013
Termination Date
May 12, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
527 units

Product Description

COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the Analyzer with or without closed tube sampling accessory A fully automated system for clinical chemistry analysis intended for the in vitro quantitative/qualitative determination of analytes in body fluids.

Reason for Recall

Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the corresponding Service Actions are not conducted automatically by the system. Additionally, certain maintenance activities that have to be performed manually by the operator are not requested on the sof

Distribution Pattern

Nationwide Distribution and Puerto Rico

Code Information

software version 9864.C2

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0168-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.