Z-0168-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- November 9, 2022
- Initiation Date
- October 3, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 63 units
Product Description
Blinq Pediatric Vision Scanner-Pediatric Vision Scanner-intended for automated detection of misalignment of the visual axes using polarized light. Model Number BQ830
Reason for Recall
Unable to charge the battery pack- battery packs will allow the batteries to be charged if the voltage of any cell is below 2.5V. When this occurs, the battery pack will not accept the charge, could vent and possibly catch fire
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of New Zealand, Belgium.
Code Information
UDI: 008648160004 Serial Numbers: B-00001-V thru B-00105-V B-00021 B-00022 B-00036 B-00052 B-00054 thru B-00060 B-00062 thru B-00065 B-00070 B-00071 B-00075 thru B-00077 B-00079 B-00082 thru B-00098 B-00100 thru B-00104 B-00106 B-00109 B-00111 thru B-00113